From my own experience: an engineer by training, I built Emobot's clinical research from scratch.
Its portfolio holds 11 clinical studies across France, the US and Canada, including 3 medical device investigations I launched in France toward CE marking, one of them across 8 sites.
Our published results show an agreement of r = 0.895 with the clinician-rated MADRS.
Is your software a medical device?
That is the first question, and the answer turns on two words: intended purpose. Regulation (EU) 2017/745 looks at the use you, the manufacturer, give your product. Software that provides information used to take decisions for diagnostic or therapeutic purposes is a medical device. Software with a non-medical purpose, such as invoicing, staff planning or a wellness app, is not. Risk does not count for qualification: it counts for the class. The software qualification guidance, MDCG 2019-11, was revised in June 2025 and also gives examples of AI software.
The trap for an engineer is to write the intended purpose like a product sheet, listing everything the software can do. Yet every word of that sentence commits your class, your studies and your timeline.
Which class, and why does it decide everything?
The regulation divides devices into classes I, IIa, IIb and III, taking into account their intended purpose and inherent risks. For software, Rule 11 of Annex VIII applies:
| Class | Software concerned (Rule 11) | Who attests conformity |
|---|---|---|
| I | Any software that fits none of the cases below | The manufacturer, through its EU declaration of conformity, once the technical documentation is drawn up |
| IIa | Information used for diagnostic or therapeutic decisions; monitoring of physiological processes | A notified body |
| IIb | Decisions that may cause a serious deterioration of health or a surgical intervention; monitoring of vital parameters whose variations could result in immediate danger | A notified body |
| III | Decisions that may cause death or an irreversible deterioration of health | A notified body |
In practice, software for diagnostic support or monitoring is at least class IIa. The class decides whether a notified body is involved, how much clinical evidence is expected, and so how long the project takes.
Where do you start, step by step?
- Write the intended purpose in one sentence. For whom, for which condition, for which decision. It sets the qualification and the class.
- Qualify and classify in writing. A short rationale that cites Rule 11 and the MDCG 2019-11 guidance. You will reread it at every step, and so will the notified body.
- Write the clinical evaluation plan. The regulation sets its minimum content, including the safety and performance requirements to support with clinical data, the intended purpose and the target groups. Conformity must rest on clinical data that provide sufficient clinical evidence, and it is up to you to justify the level of evidence needed.
- Choose the first study. The one your next step depends on, regulatory or commercial, not the most ambitious one. The plan sets which studies to run, in what order and with which endpoints.
- Write the protocol and prepare the submissions. In France, research involving people falls into one of three categories, by risk. Category 1, with non-negligible risks, needs ANSM authorization, as do trials of medical devices that do not yet have CE marking. Every trial also needs a favorable opinion from an ethics committee (CPP). Finally, plan the data protection side with the CNIL: at Emobot, I prepared it under the MR-001 reference methodology.
- Start the study. Select the sites, recruit the investigators, enroll patients on schedule. That is often where delays happen.
- Assemble the CE marking file. The clinical evaluation plan, the technical documentation, patient and healthcare professional instructions, the risk analysis, cybersecurity documentation, the post-market surveillance plan and vigilance procedures. I prepared each of these files at Emobot.
Which mistakes cost months?
- An intended purpose that is too broad, which raises the class without adding anything to the product.
- A first study chosen for its size rather than for the step it unlocks.
- A protocol mistake: in front of the ethics committee, it can cost months.
- A file built for Europe only, when you also target the US, where the FDA has its own pathways. I cover them in how to enter the US healthcare market.
When should you get help?
When your device works but you have never run a clinical study or built a regulatory file. That is where most engineer founders stand. An engineer myself, I built all of this from scratch at Emobot, where I designed the regulatory strategy and prepared the CE marking files. With my clinical research and regulatory service, I help you make the right calls from the start, from qualification to the file.